AIR MONITOR-GAS, GROSS BETA GAMMA
PARTICULATE AND IODINE
Model # BAM-4HVM
DESCRIPTION:
The BAM-4HVM series are
four channel air monitors for simultaneous measuring of gross beta/gamma
particulates and gaseous radioactivity and Iodine. Like all of the
BAM-4 series, these are continuous duty, high capacity, rugged skid
or caster-mounted systems. Electronics are microprocessor with color
LCD display. Also preamps etc. are plug-in modules allowing change
or addition of function as needed and allowing rapid repair by substitution
of modules in the field. The modular system is covered by T/A's unique
exchange warranty, in addition to the full one year warranty covering
all T/A products.
Shields are void free lead encased in welded steel
and with stainless steel liners for a long, useful life and easy decontamination.
Filters are easily changed via quick disconnect, o-ring sealed filter
holder. Shield can be opened for filter change or cleaning with minimum
effort. All shield connections and openings are sealed against air
leaks. The air moving system is based on a high capacity air pump
capable of delivering 5 cfm (Optional up
to 30 cfm) free air and including fine regulation in the 3 cfm to
5 cfm operating range. The entire system is mounted in a self-contained
rugged cabinet and comes complete with all cabling and connectors
in place ready to operate. The standard unit is operated from 115
V single phase, 60 Hz; 220 V optional, 50 or 60 Hz.
|
Figure 1: BAM-4HVM
|
SPECIFICATIONS:
- Channels: Four
- Activity Detected: Gross beta-gamma particulate,
iodine/noble gas low and high.
- Detectors:
Particulate Beta-gamma: Solid state silicon
or beta scintillation or 2'' Pancake GM Model T-1190.
Iodine: 2'' D x 2'' Nal (Tl) Model PGS3LI
spectroscopic grade.
Gas Standard Range: One Beta Scintillator
or 2 ea. GM Model T-1190.
Gas High Range: One each GM Model T-1216
- Shielding: 3'' lead; full 4 pi
- Detector Sensitivity: Monitor is more
sensitive on all channels by factors of 5 to 500 than is required
by USNRC and other regulatory agencies.
TA TECHNICAL
ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE
(818) 883-7043 * FAX(818) 883-6103
e-mail:
tagold@nwc.net www.tech-associates.com
AIR MONITOR-GAS, GROSS BETA GAMMA
PARTICULATE AND IODINE
Model # BAM-4HVM
SPECIFICATIONS (Continued):
- Particulate Beta-Gamma: 6 x 10-10
µCi/ml/min
- Iodine: 3 x 10-9 µCi/ml/min
- Gas-Standard Range: 6 x 10-7 µCi/ml/min
- Gas-High Range: High limit 1µCi/ml (Note 2)
- Display Mode: Model # BAM-4HVM features
micro-processor control and color LCD display.
- Operation: When BAM-4HVM continuous air
monitor is in normal operation, the color LCD displays:
CHANNEL NAME STATUS |
REAL TIME AIRBORNE CONCENTRATION µCi/m3 (NOTE 1) |
ACCUMULATED DOSE FOR THIS SHIFT µCi-Hrs/m3 (NOTE 1) |
CONCENTRATION ALARM STATUS |
ACCUMULATED DOSE ALARM DOSE |
BETA |
|
|
|
|
PARTICULATE |
-- |
-- |
-- |
-- |
IODINE |
-- |
-- |
-- |
-- |
NOBLE GAS |
-- |
-- |
-- |
-- |
Other alarms: |
status |
|
|
|
|
Low Air Flow ___ |
Loss of Signal Beta |
___ |
|
|
Loss of Signal Iodine |
___ |
|
|
Loss of Signal Noble Gas |
___ |
|
|
|
|
MENU OPTIONS: |
1. Modify shift length |
2. Change filter |
|
|
3. Configure energies |
4. Set alarms |
|
|
5. Start shift |
6. Save to floppy |
|
NOTE 1 - User can easily change engineering units
at will.
NOTE 2 - High Range Gas can be adjusted to see >
10µCi/ml.
Clear audio and visual alarms are provided.
- Color LCD Readout: Showing both realtime
concentration and accumulated dose for all 4 channels
- Accuracy: ± 10% except for
resolution loss and statistical variation.
TA TECHNICAL
ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE
(818) 883-7043 * FAX(818) 883-6103
e-mail:
tagold@nwc.net www.tech-associates.com
AIR MONITOR-GAS, GROSS BETA GAMMA
PARTICULATE AND IODINE
Model # BAM-4HVM
MAIN DISPLAY SCREEN
Model shown in photo is 7-channel version
DATA.TXT DISPLAY SCREEN SHOWS DATA SAVED TO FLOPPY
Model shown in photo is 7-channel version
DATA REVIEW MODE: (DATA.TXT)
Model BAM-4HVM continuous
air monitor automatically saves all data and set-up parameters to
3.5'' floppy diskette every hour. This data can e reviewed on the
CRT display.
TEST.EXE DISPLAY SCREEN FOR SETTING ENERGIES
Model shown in photo is 7-channel version
TA TECHNICAL
ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE
(818) 883-7043 * FAX(818) 883-6103
e-mail:
tagold@nwc.net www.tech-associates.com
AIR MONITOR-GAS, GROSS BETA GAMMA
PARTICULATE AND IODINE
Model # BAM-4HVM
SPECIFICATIONS (continued):
- Alarms:
Alarms presettable to any trigger
level.
Low Level Alarm: Automatic Reset
High Level Alarm: High level alarm remains activated until
RESET is pushed.
Alarm Indication: Red-High Level, Amber-Low Level
Audio: 1000 Hz, greater than 80 db. Also see Safety Features
below.
- Air Moving System:
Air pump: Extremely
rugged continuous duty Positive displacement low rotating speed pumping
system. Pump is high capacity positive displacement. Pump noise is
reduced by heavy duty muffler.
Pump Drive Motor: 1 HP, 50/60 Hz, capacitor start, induction
run.
Flow Gage: Magnehelic 3-10 cfm, 4'' dial.
Flow Control: Automatic flow control settable 0-4 cfm ±0.2 cfm.
Inlet and Outlet: 20 foot 3/4'' hose inlet: 20 foot 1''
hose outlet. May be extended.
Flow Rate: Factory set at 4 cfm, can be reset up to 10 cfm
with larger motor and up to 30 cfm with some pluming change.
- Controls:
System: ON-OFF |
Air: ON-OFF |
Printer/Recorder: ON-OFF |
(each channel) High Level Alarm: Set |
(each channel) High Level Alarm: Reset |
(each channel) Low Level Alarm:Set |
(each detector) Voltage: Set |
(each detector) Energy Threshold/Window Set |
- Outputs: Alarm condition outputs:
1. logic level signal
2. contact closure-or opening.
3. Ethernet LAN connection.
- Standard Installation: Cabinet of heavy
gage steel. Skid or caster mounted with 4 five inch ball bearing casters,
two of which are full swiveling.
- Special Safety: Loss of air flow alarm.
Loss of signal alarm. All circuits fused.
- Features: Thermal protection on motor.
Positive displacement. (Optional ROOTS
pump).
- Options: Air monitor available without
air mover or with alternate air movers. Broader range 12 decades.
Different detector types. Three push-button check sources (License
exempt quantities).
- Dimensions: 29'' wide x 31'' long x
41'' high exclusive of casters.
- Weight: 1800 lbs.
- Shipping Weight: 2100 lbs.
TA TECHNICAL
ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE
(818) 883-7043 * FAX(818) 883-6103
e-mail:
tagold@nwc.net www.tech-associates.com
AIR MONITOR-GAS, GROSS BETA GAMMA
PARTICULATE AND IODINE
Model # BAM-4HVM
ELECTRONIC SYSTEM DESCRIPTION -
Micro-processor is Pentium-based, including:
CPU with data acquisition |
Keyboard |
High resolution, color |
Color LCD display |
Mouse |
Audio |
Manuals |
Graphics printer for hard copy - 4 trace graphs |
3.5'' floppy disk - For data storage - 30 days per disk.
Plus CD Read-Write Drive.
50 gigabyte hard drive - Stores programs, plus room to add
additional new software functions in the future.
Specialized air monitor software.
Technical Associates standard modular series - all solid state.
4 ea. MGA-5202 - Fully adjustable single channel analyzer
with independent High Voltage power supply;
1 ea. SUB-1 Preamp for solid state silicon detector.
1 ea. MVR-5R Power Supply.
1 ea. LNM-5 Strip Chart Recorder or GRP-10 graphics printer.
1 ea. Magnehelic gage.
4 ea. 5202 HV bias supply/amp disc (SCA).
TA TECHNICAL
ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE
(818) 883-7043 * FAX(818) 883-6103
e-mail:
tagold@nwc.net www.tech-associates.com