BETA PARTICULATE AIR MONITOR
Model # FM-5M2
FEATURES:
- QUICK CHANGE FILTER
- REAL TIME
- LOCAL OR REMOTE SAMPLING
- INTEGRATED EXPOSURE CHART RECORDER
- CONTINUOUS OPERATION
- COMPLETELY MODULAR DESIGN
- FOUR Pi LEAD SHIELDING 2'' THICK
- REGULATED AIR FLOW
- WITH OPTIONAL PUMP AP-2 OR AP-4
APPLICATION: The FM-5M2
Air Monitor is an ideal device to assure safety against airborne
radioactive contamination even in high background or varying background
situations, by means of constant check with alarm and record capability.
It provides integrated exposure information and has digital output
available. It is a complete system but may be expanded per need with
modules of the FM-7 series.
|
Figure: FM-5M2
|
DESCRIPTION: The FM-5M2
is an all modular system which uses standard 47-55mm air filters
held in unique quick-change holder and two detector tubes contained
in a lead shielded chamber to collect and measure beta particulates
in the air. One detector measures the beta contamination while the
second detector is used for background subtraction to correct the
FM-5M2 response for changes in ambient Gamma radiation. Air flow is
regulated and adjustable and is supplied by the optional
AP-2 or AP-4 vacuum pump. The FM-5M2 alarm features include
and audible alarm and red rotating beacon as well as low level alarm.
The FM-5M2 has an auxiliary provision for remoting the alarm functions.
Time is recorded by an odometer type timer wired into the
pump circuit. Air is drawn through a three-foot inlet hose and exhausted
via a flow meter through another three-foot hose. These may be extended
to allow monitoring of hoods, glove boxes, stacks, etc... The unit
is completely self-contained and is transportable.
SPECIFICATIONS:
- Sensitivity: Sensativity is 1 x 10-12 µCi/cc Sr90Y90 after 4 hours,
at 60 L/min; 10-4 µCi per week (as beta over
0.1MeV) integrated content of entire amount of air drawn through system.
This corresponds to 10-8 µCi/ml minute exposure
which is better than 1% of California Safety Code Title 17 and NRC
Title 10 limit for 40 hour exposure to most hazardous beta and/or
gamma emitter in absence of Ac277.
- Counting Efficiency: Approximate efficiency
for 47mm diameter standard plated sources 25 percent of 2 Pi for
beta over 0.1MeV (including Tc99) and 50% of 2 Pi for Sr90Y90.
TA TECHNICAL ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE (818) 883-7043 * FAX(818) 883-6103
e-mail: tagold@nwc.net www.tech-associates.com
BETA PARTICULATE AIR MONITOR
Model # FM-5M2
SPECIFICATIONS (Continued):
- Detectors: Two each thin window 1 3/4''
x 1.4mg/cm2 pancake GM detectors Model T-1190.
- Natural Background Response:
With Background subtract turned off, typically 25cpm with
clean filter installed.
- Background Subtraction: Compensates for
approximately 90% of Gamma background. Typical background count at
0.02mR/hr is 2-4cpm.
- Shielding: 2'' (5.1cm) lead shield, in
all directions 4 Pi.
- Ranges: Four decades: 10-100,000cpm.
- Response Time: Logarithmic, varies with
count rate.
- Modules Include:
MVR-5R: Power Supply Module.
MLB-5Q: Alarming Logarithmic Ratemeter with Background Subtract.
MRR-5L: Recorder - 2 1/4'' chart.
MAC-5: Air Control Timer with Airflow Meter.
MAR-6: Alarm Relay Module.
MFT-5: Filter - Detector Module.
- Filters: Includes package of 100 filters
#41, 50mm, can use membrane paper or glass fibre filters.
- Temperature Range: 20 °F
to 120°F (-7°C to 49°C)
with less than ±20% change in calibration and alarm
point.
- Finish: Baked Enamel, easily cleaned.
- Dimensions: Excluding bell and beacon: 17'' W x 18'' H x 14 1/2'' D.
- Weight: 110lbs.
- Shipping Weight: 130lbs.
- Power: 115V, 60Hz (230V, 50Hz Optional).
Optional Pumps:
Model: |
AP-4 |
AP-2 |
Dimensions: |
7'' W x 16'' H x 28'' L |
6'' W x 8'' H x 13'' L |
Weight: |
24 lbs. |
20 lbs. |
Flow Rate: |
Regulated ±5% from near zero to maximum pump capacity 120 liters/min (4cfm). |
Adjustable 30 liter/min (1cfm) |
Pump Type: |
Oilless, Carbon Vane or Equivalent. |
Oilless, Carbon Vane |
Motor: |
115V, 60Hz, 6 amps. |
115V, 60Hz, 2 amps. |
TA TECHNICAL ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE (818) 883-7043 * FAX(818) 883-6103