MANUAL LIQUID SCINTILLATION
COUNTING SYSTEM
Model # SSS-22
FEATURES:
- HIGH SENSATIVITY H3, C14, ALL
BETA
EMITTERS AND LOW ENERGY GAMMA EMITTERS
- DUAL PM TUBE DESIGN
APPLICATION: The SSS-22
Manual Liquid Scintillation Counting System accurately quantitatively
measure Carbon-14, Tritium and other radioactive labels.
SYSTEM DESCRIPTION:
|
Figure 1: SSS-22
|
Measuring Priciple: The most sensative
method of detecting and quantitating beta emitting isotopes is to
intimately mix the sample with liquid scintillation fluor and count
each individual scintillation event with a dual photomultiplier coincidence
counter. Followed by an energy analyzer which further selects the
pulses and delivers the true signal. Hard copy printer is optional.
Detection cell optically coupled to two selected photomultiplier
tubes which count in coincidence mode. Coincidence counting eliminates
photo-cathode noise pulses, that are always present in any one-PM-tube
system. Extremely good figure of merit for low energy betas such as
H3 and C14 are thus achieved.
GAMMA BACKGROUND RADIATION REJECTION FEATURES:
- Lead shielding around detector
- Energy Analyzing window rejects pulses with energies
outside the window setting
COUNTING ASSEMBLY FEATURES:
- Excellent repeatability
- Fully light tight system
- Fail safe interlock to protect PM tubes
PM TUBE AND PRE-AMP NOISE IS ELIMINATED BY THESE FEATURES:
- Coincidence counting - dual PM Tubes
- High quality PM tubes and preamps
- Fully adjustable energy analyzer window rejects
low energy pulse
Settable window can be set for any isotope. C14 window can be set for maximum responce, or can be set to exclude Tritium.
READOUT IN PERCENT OR IN COUNTS
Preset count and preset time functions allow direct
comparison of standard and sample: The SSS-22 allows
readout of sample activity directly as a percent of the standard,
in counts/sec or counts/minute. Optionally, gamma ray counting is
achieved by inserting and optically coupling an Na(Tl) scintillation
well crystal between the PM tubes.
TA TECHNICAL
ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE
(818) 883-7043 * FAX(818) 883-6103
e-mail:
tagold@nwc.net www.tech-associates.com
MANUAL LIQUID SCINTILLATION
COUNTING SYSTEM
Model # SSS-22
DATA ANALYSIS AND PRESENTATION
Scintillation counts which are detected simultaniously
by both PM tubes are processed by a fully adjustable single channel
analyzer which is centered on the energy peak of the isotope being
measured. This deletes both higher energy pulses from background radiation
and lower energy counts from the PM tube or circuit noise. The pulses
are then fed to a digital scaler and Optional digital
printer. (Thus allows long count times for measurement of very minute
samples as well as completely elimination artifacts caused by ratemeter
time constants.) Built-in RS-232 serial input of most computers.
Optional Win-wedge data logging software.
SPECIFICATIONS:
- Tritium Efficientcy: In excess of 60%
for all systems.
- C-14 Efficientcy: Typically 90%.
- For Dual Channel System: C14 Efficiency: 15% when H3 window is set for 50% H3 efficiency.
- Count Times: 1 sec thru 99,999 sec, in
1 sec increments.
- Voltage: 0-2000 Volts - 2 fully independant
HV supplies, individually settable.
- Readout: Digital - 5 digit plus overflow.
- Power: 110VAC, 60 Hz (220V 50 Hz Optional).
- Dimensions: Electronics console 21''W
x 18'' x 11''H. Detector enclosure 18'' x 6'' x 6''.
- Weight: 45lbs.
- Shipping Weight: 65lbs.
- Sample Size: Accepts standard Liquid
Scintillation vials up to 1.1'' dia. x 2 1/2''H.
- Scintillation Fluors: Accepts most scintillation
fluors.
- Options: Digital Printer Model MPM-5.
Gamma Well Counter Crystal, NaI(Tl).
Detector for SSS-22