MANUAL LIQUID SCINTILLATION
COUNTING SYSTEM

Model # SSS-22

FEATURES:
  • HIGH SENSATIVITY H3, C14, ALL BETA
    EMITTERS AND LOW ENERGY GAMMA EMITTERS
  • DUAL PM TUBE DESIGN

APPLICATION: The SSS-22 Manual Liquid Scintillation Counting System accurately quantitatively measure Carbon-14, Tritium and other radioactive labels.
 
SYSTEM DESCRIPTION:
 
SSS-22 Image

Figure 1: SSS-22

Measuring Priciple: The most sensative method of detecting and quantitating beta emitting isotopes is to intimately mix the sample with liquid scintillation fluor and count each individual scintillation event with a dual photomultiplier coincidence counter. Followed by an energy analyzer which further selects the pulses and delivers the true signal. Hard copy printer is optional.

Detection cell optically coupled to two selected photomultiplier tubes which count in coincidence mode. Coincidence counting eliminates photo-cathode noise pulses, that are always present in any one-PM-tube system. Extremely good figure of merit for low energy betas such as H3 and C14 are thus achieved.

GAMMA BACKGROUND RADIATION REJECTION FEATURES:


COUNTING ASSEMBLY FEATURES:

PM TUBE AND PRE-AMP NOISE IS ELIMINATED BY THESE FEATURES:
Settable window can be set for any isotope. C14 window can be set for maximum responce, or can be set to exclude Tritium.
 
READOUT IN PERCENT OR IN COUNTS

Preset count and preset time functions allow direct comparison of standard and sample: The SSS-22 allows readout of sample activity directly as a percent of the standard, in counts/sec or counts/minute. Optionally, gamma ray counting is achieved by inserting and optically coupling an Na(Tl) scintillation well crystal between the PM tubes.


TA  TECHNICAL ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE (818) 883-7043 * FAX(818) 883-6103
e-mail: tagold@nwc.net     www.tech-associates.com


MANUAL LIQUID SCINTILLATION
COUNTING SYSTEM

Model # SSS-22

DATA ANALYSIS AND PRESENTATION

Scintillation counts which are detected simultaniously by both PM tubes are processed by a fully adjustable single channel analyzer which is centered on the energy peak of the isotope being measured. This deletes both higher energy pulses from background radiation and lower energy counts from the PM tube or circuit noise. The pulses are then fed to a digital scaler and Optional digital printer. (Thus allows long count times for measurement of very minute samples as well as completely elimination artifacts caused by ratemeter time constants.) Built-in RS-232 serial input of most computers. Optional Win-wedge data logging software.

SPECIFICATIONS:


SSS-22 Detector

Detector for SSS-22



TA  TECHNICAL ASSOCIATES
7051 ETON AVENUE * CANOGA PARK, CA 91303
TELEPHONE (818) 883-7043 * FAX(818) 883-6103
e-mail: tagold@nwc.net     www.tech-associates.com