TA RADON MONITOR COMPARISION CHART





Feature                                                            PTG-7RN                                                TBM-IC-RN


Lower level of detection                               5 x 10-11 uC1/ml                                   4 x 10-9 uC1/ml
                                                                         (2 Bq/m3)                                           (4PCi/l)(148 Bq/m3)
Measurement time required                               1 hour                                                      10 minutes
Accuracy of mid-range measurement        +/- 10% or better                                    +/- 20% or better 

Lower level of detection                                  2 x 10-9 uC1/ml                                   4 x 10-9 uC1/ml  
                                                                            (74 Bq/m3)                                        (4pCi/l) (148 Bq/m3)
Measurement time required                                1 minute                                                10 minutes
Accuracy of mid-range measurement          +/- 10% or better                                       +/- 20%


Lower level of detection                                 1 x 10-8 uC1/ml                                       1 x 10-8 uC1/ml  
                                                                        (1000 Bq/m3)                                    (10pCi/l) (1000 Bq/m3)
Measurement time required                             5 seconds                                                  5 seconds
Accuracy of mid-range measurement             +/- 10%                                                      +/- 40%


TECHNOLOGY                                                 PTG-7RN                                                    TBM-IC-RN

Sensor                                                            Flow-thru Ion Chamber                        Flow-thru Ion Chamber
Volume                                                             2000 milliLiters                                           300 milliLiters
Background chamber                                            Yes                                                            No
Portable                                                                   Yes                                                           Yes
Fixed use                                                                 OK                                                            No
Power                                                               Battery or AC                                             Battery  Only
Display                                                               6 digit LED                                                6 digit LED
Alarm                                                                       Yes                                                            No
Data output                                                      Serial RS-232                                                 No
Pump                                                                    Built-in                                                      Optional
Draws Air from                                             Glove-box, Smoke-stack                             Room-air
                                                                            Or Room-air                                        Optional "sniffer hose"